1-15
IM 04L51B01-01EN
Configuring the GX/GP
1
Moving average
Setup Item Selectable Range or Options Default Value
On/Off Off, On Off
Count
1
2 to 100 2
1 Appears when On/Off is set to On.
On/Off
Set this to On to perform moving average.
Count
Set the number of data points to take the moving average of.
Average
Scan interval
Sampling count
Average
RJC
1
(Reference junction compensation)
Setup Item Selectable Range or Options Default Value
Mode Internal/External Internal
Temperature
2
–20.0 to 80.0°C, 253.1 to 353.2K (KpvsAu7Fe) (when the
measurement range is in Fahrenheit: –4.0 to 176.0, 253.1
to 353.2K)
0.0°C,
273.2K (KpvsAu7Fe)
1 Appears when the range type is set to TC.
2 Appears when the mode is set to External.
Mode
Set the reference junction compensation method of the thermocouple.
Options Description
Internal Uses the reference junction compensation function of the GX/GP.
External Uses an external reference junction compensation function.
External reference junction
Internal reference junction
Thermocouple
Thermocouple
Temperature
When the RJC is set to external, set the compensation temperature.
1.2 Configuring AI Channels (Analog input (including DI) channels)