IM DX1K-S2E Ed-01 User’s manual
9 2.10. Communication Commands 2.10.1. Setting the LOG Range Setting mode setting SR p1,p2,p3,p4,p5,p6,p7,p8,p9 <terminator> ¾ p1: Chan
10 3. Accuracy and Display Resolution 3.1. Measurement and Display Accuracy (Digital Display) Computing method of the display value accuracy ¾
11 Computing example of the display value accuracy ¾ Integration time 50 Hz ¾ 2 V range Voltage span 0.0000 to 1.0000 V ¾ LOG span -2 to 3 1
12 4. Simulation Function 4.1. Display Freeze Function 4.1.1. Specifications A function used to stop the updating of the time, trend, digital v
13 Event action functions that do not operate while the display is frozen ¾ See Table 4-1. ¾ If a level action such as Memory start/stop is execut
14 4.1.3. Setting the Display Freeze Function 1. Press MENU (switch to the setting mode). 2. Select Timer, Event action > Event action. 3. Set th
15 4.2. Memory and Trend Clear Function 4.2.1. Specifications A function used to clear the internal memory and the trend. ¾ Clears the data area
16 4.2.3. Setting the Memory and Trend Clear Function 1. Press MENU (switch to the setting mode). 2. Select Timer, Event action > Event action. 3
17 4.4. Enabling/Disabling the Time Display 4.4.1. Specifications A function used to disable the time display on the operation display. Display
18 4.4.2. Enabling/Disabling the Time Display 1. Press MENU (switch to the setting mode). 2. Hold down FUNC for 3 s to switch to the basic setting
1 Contents 1. Foreword ...2 2. L
19 4.5. Special Current Value Mark Function 4.5.1. Specifications A function used to display the channel using two digits for the current value
20 4.5.2. Setting the Channel Display of the Current Value Mark 1. Press MENU (switch to the setting mode). 2. Hold down FUNC for 3 s to switch to t
2 1. Foreword Thank you for purchasing the DX1000(N)/DX2000 DXAdvanced. This user’s manual describes the functions listed below. For a description o
3 2. LOG Input Specifications 2.1. Converting the LOG Input (Relationship between the Input Voltage and Digital Display) Trend display (logarithm
4 2.2. LOG Input Mode This mode allows you to measure the DC voltage input (DCV: 20 mV, 60 mV, 200 mV, 2 V, 6 V, 20 V, and 50 V), convert to logar
5 Table 2-1 Log Input Mode Settings Setup Item Selectable Items (Selectable Range) Description Range Same as the DCV range Sets the range. Span Low
6 2.3. Alarm Alarm types in LOG input mode ¾ H, L, T, and t only. Set the alarm value using a voltage. Computing method of alarm values ¾ Co
7 2.4. Computation Function (Option) When a channel set to LOG is used in a computation, error data is used as its measured value. 2.5. Report
8 2.8. Communication Output and Binary Data Output Format The binary data is an A/D normalized value of the channel set to LOG. ¾ Measured data