<Appendix2. ILBD Check List>
A2-2
IM 01C25T01-06EN
Checklist (2/5)
No. Items Parameters Result Example
8
Alarm status
Check the alarm status shown in Diag Error.
Check that the alarm status of “ILBD over
range” is not shown in Diag Error.
Diag Error
A Blocking
Large Fluct L
Large Fluct H
L Side Blocking
H Side Blocking
B Blocking
Invalid Ref F
Invalid Ref SPH
Invalid Ref SPL
Invalid Ref DP
ILBD over range
9
ILBD parameters
Record the values of parameters for ILBD
operation.
Check the status of parameters for ILBD
operation.
*: Record the value after checked that the
status of each parameter is “GOOD”.
Diag Period 180
Lim fDPmax 3.000000
Lim fDPmin 0.300000
Lim fSPlmax 5.000000
Lim fSPlmin 0.500000
Lim fSPhmax 5.000000
Lim fSPhmin 0.500000
Lim BlkFmax 0.600000
Lim BlkFmin -0.600000
Lim DPAvgmax 1.000000
Lim DPAvgmin 0.050000
Diag Supp Count 3
Ref fDP* 7.43245E-09
Ref fSPl* 7.25765E-09
Ref fSPh* 7.18374E-09
Ref DPAvg* 5.36425E+00
fDP* 7.48562E-09
fSPl* 7.23277E-09
fSPh* 7.14085E-09
BlkF* -0.287259
DPAvg* 0.055957