<3. Parameter Setting>
3-32
IM 01C25T01-06EN
b) Acknowledge Event Notication
Execute Acknowledge Event Notication
method.
1) Set to Enter Event Number is in conrmed
Event Number a)5.
2) OK.
3) Set Trans 1: Send Acknowledge” to Select
Transaction.
4) OK.
5) Conrm Event Status is 0x00.
(3) Event Notication Record
Procedure to call up the display
DD and DTM [Root Menu] Detailed setup
Output condition →HART output
Burst condition Event
Notication Knowledge
Acknowledge
Event Notication
Acquisition of the event number
and approval.
a) Get Event Number
Conrm the latest event number.
Execute Acknowledge Event Notication
method.
1) Enter Event Number is set to “0”.
2) OK.
3) Set Trans 0: Read Event Noticaiton” to
Select Transaction.
4) OK.
5) Conrm Event Number.
b) Conrmation record of Event Notication
Conrm four events checked in a).
Execute Acknowledge Event Notication
method.
1) Enter the event number to Enter Event
Number which is conrmed in a)5.
2) OK.
3) Set Trans 0: Read Event Noticaiton” to
Select Transaction.
4) OK.
5) Knowledge menu displays events record.
Ex.) When the conrmed event number is 123.
Event Number Explanation
123 The latest event
122 An event before the once.
121 An event before the twice.
120 An event before three times.
3.3.17 Multidrop Mode
3.3.17.1 Setting on HART 5
HART 5
“Multidropping” transmitters refer to the connection
of several transmitters to a single communication
transmission line. Up to 15 transmitters can be
connected when set in the multidrop mode. To
activate multidrop communication, the transmitter
address must be changed to a number from 1 to
15. This change deactivates the 4 to 20 mA analog
output, sending it to 4 mA. The alarm current is also
disabled.
Setting of Multidrop Mode
(1) Polling address
Procedure to call up the display
DD and DTM
(excluding EJX_
HART 5[1.2])
[Root Menu] Detailed setup
Output condition HART output
EJX_HART 5[1.2]
DTM
Conguration HART
Poll addr Enter the number from 1 to 15
(2) Enabling the Multidrop Mode of
Conguration Tool
About the procedure to call up the Polling
display, please refer to the Users Manual of each
conguration tool.
NOTE
When the same polling address is set for
two or more transmitters in multidrop mode,
communication with these transmitters is
disabled.
(3) Communication when set in multidrop
mode.
The HART conguration tool searches for a
transmitter that is set in multidrop mode when
it is turned on. When the HART conguration
tool is connected to the transmitter, the polling
address and the tag will be displayed.
Select the desired transmitter. After that, normal
communication to the selected transmitter is
possible. However, the communication speed
will be slow.
To release multidrop mode, call up the Poll addr
display and set the address to “0”.