1. CAUTIONARY NOTES FOR SAFE
USE OF THE PRODUCT
For the correct use of this product, read through this manual before
use. The following safety symbol is indicated on the product to en-
sure safe use.
CAUTION
If this symbol is indicated on the product, the operator
should refer to the explanation given in the instruction
manual in order to avoid personnel injury or death to either
themselves or other personnel, and/or damage to the instru-
ment. The manual describes the special care the operator
should exercise to avoid shock or other dangers that may re-
sult in injury or loss of life.
The following symbol marks are used only in this manual.
IMPORTANT
Indicates that operating the hardware or software in a par-
ticular manner may damage it or result in a system failure.
NOTE
Draws attention to information that is essential for under-
standing the operations and/or features of the product.
2. CHECKING PRODUCT SPECIFI-
CATIONS AND THE CONTENTS
OF PACKING
(1) Model Number and Specification Check
Check that the model number and specifications shown on the
nameplate attached on the side of the product are as ordered.
(2) Contents of the Packing
Check that the packing contains the following items:
VJA7 main unit, 1
Instruction Manual (IM 77J1A07-01E), 1
Tag number label, 1
3. GENERAL
This distributor, which belongs to the JUXTA series of signal condi-
tioners, receives 4 to 20 mA DC signals from two-wire transmitters
while supplying power to transmitters and converting the received
signal to pairs of isolated current and voltage signals.
Output-2 signal is selectable from a DC voltage signal, DC
current signal, communication function (RS-485), and alarm
output (two relay contacts).
4. MOUNTING METHODS
4.1 Wall Mounting
Loosen the main unit-fixing screw of the distributor to disconnect
the main unit from the socket. Next, anchor the socket onto the
wall with screws. Then, plug the main unit into the socket and se-
cure the main unit with the main unit-fixing screw.
Main
unit-fixing
screw
Socket
Main unit
Threaded hole for
fixing the main unit
Mounting
screws
<Mounting Dimensions>
Unit: mm
29.5 or more
22±0.2
59±0.3
2-M4 or 2-ø4.5 or more
Fig. 4.1
4.2 DIN Rail Mounting
Locate the distributor so that the DIN rail fits into the upper part of
the DIN-rail groove at the rear of the socket, and fasten the socket
using the slide lock at the lower part of the socket.
Slide lock
DIN rail
DIN rail
VIEW A (Rear of socket)
VIEW A
Fig. 4.2
Keep this manual in a safe place.
Network Solutions Business Divisiion
2-9-32, Naka-cho Musashino-shi, Tokyo 180-8750 Japan
Phone: +81-422-52-7179 Facsimile: +81-422-52-6793
Page view 0
1 2 3 4 5

Summary of Contents

Page 1 - Distributor (Multi-function)

User’sManualModel VJA7Distributor (Multi-function)(Isolated Single-output andIsolated Dual-output Types)IM 77J01A07-01EIM 77J01A07-01E1st Edition Nov.

Page 2 - 6. EXTERNAL WIRING

2All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation IM 77J01A07-01E 3rd Edition Mar.31,1997-004.3 Mounting Using a Multi-mounting Ba

Page 3 - 8. SETTING PARAMETERS

3All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation IM 77J01A07-01E 3rd Edition Mar.31,1997-00IMPORTANT● If this instrument is used

Page 4 - 8.3.3 Hysteresis

4All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation IM 77J01A07-01E8.3.2 Direction of Alarm ActionSelect the direction of alarm-1 act

Page 5 - 11.MAINTENANCE

5All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation IM 77J01A07-01E10.LIST OF PARAMETERS.oNmetIyalpsiDskrameR.oNmetIyalpsiDskrameR10l