Yokogawa JUXTA VJU7 User Manual

Browse online or download User Manual for Sensors Yokogawa JUXTA VJU7. Yokogawa JUXTA VJU7 User Manual [en] [es] [fr]

  • Download
  • Add to my manuals
  • Print
  • Page
    / 8
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Users
Manual
Model VJU7
Universal Temperature Converter
(Isolated Single-output and
Isolated Dual-output Types)
IM 77J01U07-01E
4. MODEL AND SUFFIX CODES
0/
Input signal
U: Thermocouple, RTD, mV signal
Output-1 signal
A: 4 to 20 mA DC
6: 1 to 5 V DC
Z: (Custom Order) DC current or voltage signal
Output-2 signal
A: 4 to 20 mA DC
6: 1 to 5 V DC
P: Communication function (RS-485)
T: Alarm output (2 relay contacts)
N: No output-2 signal
Power supply
6: 100-240 V AC/DC
(Operating range: 85 to 264 V)
7: 15-30 V DC (Operating range: 12 to 36 V)
Number of outputs
1: 1 output
2: 2 outputs
Model
VJU7-0 -
Optional specification
Blank: With socket
/
SN: Without socket
/
RJCN: Without RJC sensor
1. CAUTIONARY NOTES FOR SAFE
USE OF THE PRODUCT
This user’s manual should be carefully read before installing
and operating the product. The following symbol is used on the
product and in this manual to eusure safe use.
WARNING
This symbol is displayed on the product when it is
necessary to refer to the users manual for infor-
mation on personnel and instrument safety. This
symbol is displayed in the users manual to indi-
cate precautions for avoiding danger to the opera-
tor, such as an electric shock.
The following symbols are used only in this manual.
IMPORTANT
Indicates that operating the hardware or software
in a particular manner may cause damage or result
in a system failure.
NOTE
Draws attention to essential information for under-
standing the operations and/or functions of the
product.
2.
CHECKING PRODUCT SPECIFICATIONS
AND PACKAGED ITEMS
(1) Model and Specifications Check
Check that the model and specifications indicated on the
nameplate attached to the side face of the main unit are as
ordered.
(2) Contents of the Package
Check that the package contains the following items:
VJU7: 1
User’s manual (this manual: IM 77J01U07-01E): 1
Accessories:
Tag number label: 1 sheet
Range label: 1 sheet
RJC sensor (when /RJCN is not specified for the
optional specification): 1
3. GENERAL
This plug-in type Universal Temperature Converter receives
thermocouple, RTD or mV signal and converts it into isolated
DC voltage or current signal.
Settings of the input type (thermocouple, RTD or mV)
and measurement range can be changed within the
specified ranges.
One of DC voltage signal, DC current signal, communi-
cation function (RS-485) and alarm output (2 relay con-
tacts) can be selected for the output-2 signal.
Keep this manual in a safe place.
Network Solutions Business Divisiion
2-9-32, Naka-cho Musashino-shi, Tokyo 180-8750 Japan
Phone: +81-422-52-7179 Facsimile: +81-422-52-6619
IM 77J01U07-01E
©Copyright Nov. 1999
4th Edition: Feb. 2007
Page view 0
1 2 3 4 5 6 7 8

Summary of Contents

Page 1 - Isolated Dual-output Types)

User’sManualModel VJU7Universal Temperature Converter(Isolated Single-output andIsolated Dual-output Types)IM 77J01U07-01E4. MODEL AND SUFFIX CODES0/

Page 2 - 7. EXTERNAL WIRING

2All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation IM 77J01U07-01E 4th Edition Feb. 09, 2007-005. MOUNTING METHODS5.1 Wall Moun

Page 3 - 9. SETTING PARAMETERS

3IM 77J01U07-01E 4th Edition Feb. 09, 2007-00●Use of the product ignoring the specificationsmay cause overheating or damage. Before turn-ing on the p

Page 4 - Function

4All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation IM 77J01U07-01E 4th Edition Feb. 09, 2007-009.1.2 Input Hard RangeSelect the

Page 5 - 9.3.3 Hysteresis

5IM 77J01U07-01E 4th Edition Feb. 09, 2007-009.3 Settings Related to Alarm OutputSet the following parameters when the alarm output is speci-fied

Page 6

6All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation IM 77J01U07-01E 4th Edition Feb. 09, 2007-0010. DESCRIPTION OF ALARM ACTIONSThi

Page 7 - 11. LIST OF PARAMETERS

7IM 77J01U07-01E 4th Edition Feb. 09, 2007-0011. LIST OF PARAMETERS No. Item Display 01 Model MODEL 02 Tag No. TAG NO 03 Self-check result

Page 8 - 12.2 Calibration Procedure

8All Rights Reserved. Copyright © 1999, Yokogawa M&C Corporation IM 77J01U07-01E 4th Edition Feb. 09, 2007-0012. MAINTENANCEThe product enters t

Comments to this Manuals

No comments