Yokogawa Touch Screen GP20 User Manual Page 33

  • Download
  • Add to my manuals
  • Print
  • Page
    / 110
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 32
2-19
IM 04L51B01-17EN
Commands and Responses
1
2
App
Example Set three set points on channel 0001
(measurement range: 0 to 1.0000 V). Set the
set points as follows: when the input value is 0
V, the output value is 0.0010 V; when the input
value is 0.5000 V, the output value is 0.5020
V; when the input value is 1.0000 V, the output
value is 0.9970 V.
SCalibIO,0001,Appro,3,0,10,5000,
5020,10000,9970
Description
If p2=Off, you cannot set p3 or subsequent
parameters.
You cannot specify set points beyond the number of
points specified by p3.
If the AI channel input type (p2 of SRangeAI) is set
to Skip or DI, you cannot specify anything other than
p2=Off.
SMathBasic
Math Action (/MT)
Sets the basic operation of math channels.
Syntax
SMathBasic,p1,p2,p3,p4
p1
Indication on computation error
+Over
Display the computed
value as +Over.
-Over
Display the computed
value as -Over.
p2
SUM and AVE computation when overflow
data is detected
Error
Sets the computation
result to computation
error.
Skip
Discards the data that
overflowed and continues
the computation.
Limit
Computes by substituting
upper or lower limit
values in the data that
overflowed.
For channels that do
not have linear scaling
specified, the upper
or lower limit of the
measuring range
For channels that
have linear scaling
specified, the scaling
upper or lower limit
For math channels,
the specified span
upper or lower limit.
p3
MAX, MIN, and P-P computation when
overflow data is detected
Over
Computes using data that
overflowed.
Skip
Discards the data that
overflowed and continues
the computation.
P4
START/STOP key action
Off
Computation does not
start even when recording
starts.
Start/Stop
Computation starts when
recording starts.
Reset+
Start/Stop
Computation resets and
starts when recording
starts.
Query
SMathBasic?
Example Set the indication on computation error to
“+Over,” computation when overflow data is
detected to “Skip,” and start computation when
recording starts.
SMathBasic,+Over,Skip,Skip,Start/
Stop
2.4 Setting Commands
Page view 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 109 110

Comments to this Manuals

No comments