IM 12D06D05-01E4th EditionUser’sManualModel ISC450G [Style: S2]Inductive Conductivity Converter
2 IM 12D06D05-01E2. GENERAL SPECIFICATIONS OF EXAxt ISC450GA) Input specifications :CompatiblewiththeYokogawainductiveconductivityISC40serie
3IM 12D06D05-01E2 GENERALSPECIFICATIONS Hold : ContactcanbeusedtosignaltheHOLDsituation. Failsafe : ContactS4isprogrammedasfail-sa
4IM 12D06D05-01EN) Safety and EMC conforming standards , Safety : EN61010-1 CSAC22.2No.61010-1 UL61010-1 FM3611ClassI,
5IM 12D06D05-01E3 INSTALLATIONANDWIRING3-1. Installation and dimensions3-1-1. Installation siteTheEXAxt450converterisweatherproofandcan be i
6 IM 12D06D05-01Emin.185 (7.25")min.195(7.75")138(5.43")138(5.43")M6M6M5138(5.43")+1 0138(5.43")+1 0Figure 3-2. Opt
7IM 12D06D05-01E3 INSTALLATIONANDWIRING3-2. Wiring3-2-1. PreparationRefertofigure3-4.Therelaycontactterminalsand power supply connections a
8 IM 12D06D05-01EHigh voltage sectionContact(S1, S2)outputcablesmAcablesContact(S3, S4)outputcablesSensorCablesInputcontactPowercableSuitable for cabl
9IM 12D06D05-01E3 INSTALLATIONANDWIRING3-3. Wiring the power supply3-3-1. General precautionsMake sure the power supply is switched off. Also, make
10 IM 12D06D05-01E3-3-5. Grounding the housingForthesafetyoftheuserandtoprotecttheinstrument against interference, the housing must always b
11IM 12D06D05-01E3 INSTALLATIONANDWIRING3-4. Wiring the contact signals3-4-1. General precautionsThe contact output signals consist of volt-age-fre
EnterMeasurement setup Temperature settings Temp. Compensation Calibration settings Concentration Measure Conductivity only Configure sens
12 IM 12D06D05-01EFigure 3-9a. DimensionsIt is necessary to use screening/shielding on the output signal cables. Terminal 63 is used to connect the s
13IM 12D06D05-01E3 INSTALLATIONANDWIRINGFigure 3-9b. Dimensions3-6-4. Sensor cable connections using junc-tion box (BA10) and extension cable (WF1
14IM 12D06D05-01EExtension cable may be purchased in bulk quantities or in pre-finished lengths. In the case of bulk quantities cut to length, then i
15IM 12D06D05-01E3 INSTALLATIONANDWIRINGFigure 3-11.a. Figure 3-11.b. Figure 3-11.c.
16 IM 12D06D05-01E4-1. Main display functionsGo to trend screen Go to zoom screenGo to status screenGo to maintenance screenTag :EXAxt ISC4503.76025.
17IM 12D06D05-01E4OPERATIONOFEXAxtISC450G4-3-1. Actual mA1 = the current output in mA of the first current output, which is defined as mA1. The ra
18 IM 12D06D05-01E4-3-14. HART Device revisionSometimesthefirmwareofadeviceisupdated in a way that the communication file (HART DD) need revisi
19IM 12D06D05-01E4OPERATIONOFEXAxtISC450G“RETURNKEY”exittopreviousdisplayMain displayPrimary setup displayCommisioning menu displayInstrument
20 IM 12D06D05-01E5-1. Configure sensor Measuring unit /cm /mEither /cm or /m can be chosen here. The Proc-ess values will be expressed in S/cm or S/m
21IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.ConfigureSensor Cellconstant 1.88cm-1 0.2
Electrostatic dischargeThe EXAxt converter contains devices that can be damaged by electrostatic discharge. When servicing this equipment, please obse
22 IM 12D06D05-01E5-4. Calibration settingsAir adjust limitTo avoid cable influences on the measurement, a“zero”calibrationwithadrysensormaybe
23IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.Calibration Airadjust 100.00µS 0µS 100
24IM 12D06D05-01E5-6. mA output setupThe general procedure is to first define the function (control, output, simulate, off) of the output and second
25IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.mA1 (output) Damping time 0.0 sec. 0.0 sec.
26 IM 12D06D05-01EExpire timeIftheoutputisover100%forlongerthantheexpiretime,theoutputwillreturnto0%.Damping timeThe response to a st
27IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.AlarmS1 Setpoint 900.0mS/cm(high) -inf
28 IM 12D06D05-01E5-8. FailA fail contact is energized when a fail situation occurs.Failsituationsareconfiguredinsecton5-11.ForSOFTFailsthe
29IM 12D06D05-01E5MENU STRUCTURECOMMISSIONING Menu Parameter Default Range values min. max.Simulation Percentage 50% 0% 100%S2,S3,
30 IM 12D06D05-01E5-11. Error configurationErrors 1/3 ~ 3/3Errors are intended to notify the user of any unwanted situations. The user can determine w
31IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values min. max.Errors Highlimit 1.000S 0.1S 3SErrors Lo
ii The following safety symbols are used on the product as well as in this manual.DANGERThis symbol indicates that an operator must follow the instruc
32 IM 12D06D05-01E5-13. Advanced setupDefaultsThe functionality of the EXAxt allows to save and load defaults to come to a known instru-ment setting.
33IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values Low HighHART Network address 0 0 15
34IM 12D06D05-01E5-14. Display setupMain displayThe main display consists of three lines with Process Values. Each line is user definable with the re
35IM 12D06D05-01E5MENU STRUCTURECOMMISSIONINGMenu Parameter Default Range values Low HighY-axis Conductlow 0µS/cm -inf +infY-ax
36 IM 12D06D05-01E6. CALIBRATION6-1. Before calibration1. When is calibration necessary?Calibration of con duc tiv i ty in stru ments is not nor
37IM 12D06D05-01E6CALIBRATIONCell constant: The nominal cell constant of the sensor is 1.88 cm-1forthePEEKsensortypes and 3.00 cm-1forthePFA
38 IM 12D06D05-01E6-2. Cell constant manualThe intention of this calibration routine is to fine tune a sensor for which only the nominal cell constant
39IM 12D06D05-01E6CALIBRATION6-7. Temperature calibration In order to make the most accurate measure-ments, it is important to have a precise temper-
40IM 12D06D05-01E7-1. Periodic maintenance The converter requires very little periodic main-tenance, except to make sure the front window is kept cle
41IM 12D06D05-01E7 MAINTENANCETouchscreen adjustment A few years after using, the touchscreen may deviate from the correct position due to aging de
IM 12D06D05-01EiiiIM 12D06D05-01E4th Edition: Mar. 2012 (YK) All Rights Reserved, Copyright © 2007, Yokogawa Electric CorporationTABLE OF CONTENTSPREF
42IM 12D06D05-01E8-1. General The EXAxt is a microprocessor-based analyzer that performs continuous self-diagnostics to verify that it is working cor
43IM 12D06D05-01E9QUALITYINSPECTION9. QUALITY INSPECTIONQualityInspectionStandards ISC450GInductive Conductivity Converter QIS 12D06D05-01E1st Ed
44IM 12D06D05-01E2/3 QIS 12D06D05-01EThis test is done on the “HIF” display of “Factory Mode”. a. Touch the [Setup] icon. b. Touch the [Commissioni
45IM 12D06D05-01E9QUALITYINSPECTION3/3 QIS 12D06D05-01ETable 4 Data Display Current Output (mA DC) 4 4 ±0.02 12 12 ±0.02 20 20 ±0.02 After all t
46IM 12D06D05-01E
47IM 12D06D05-01E10SPAREPARTS10. SPARE PARTSSeeCustomerMaintenancePartsList.
48IM 12D06D05-01EAPPENDIX 1Temperature compensationTheconductivityofasolutionisverydependentontemperature.Typicallyforevery1°Cchangein
49IM 12D06D05-01EAAPPENDICESB. Calculation of temperature coefficient factor(with two known conductivity values at different temperatures)Measure the
50 IM 12D06D05-01E• CheckingWhen the temperature coefficient already set is accurate, the conductivity to be displayed must be constant regardless of
51IM 12D06D05-01EAAPPENDICESThe matrix can be cleared before entering new values. Next new matrix values can be entered as described above. The EXAxt
IM 12D06D05-01Eiv 4. OPERATION OF EXAxt ISC450G ...164-1.Maindisplayfunct
52 IM 12D06D05-01EAPPENDIX 3Calibration solutions for conductivityNote! This section should be read in conjunction with the calibration section (Chapt
53IM 12D06D05-01EAAPPENDICESIf it is more convenient, the user may make solutionsfromSodiumChloride(NaClorcom-mon table salt) with the help of t
54IM 12D06D05-01EOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuProcess valuesPrimary value (SC/Conc.)Secondaryvalue(Temp.)Tertiary
55IM 12D06D05-01EAAPPENDICESOnline menu Level 1 menu Level 2 menu Level 3 menuLevel 4 menuMost appl. ErrorCalibrationError description / remedyCCCali
56 IM 12D06D05-01EOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuCommissioningOutput setup mA1 setupmA2 setup (similar to mA1)Type = c
57IM 12D06D05-01EAAPPENDICESOnline menu Level 1 menu Level 2 menu Level 3 menu Level 4 menuLoop testBasic setup TagDistributorModelDevice information
58 IM 12D06D05-01EAPPENDIX 6, Control drawing for FM approval
CMPL12D06D05-02ECustomerMaintenanceParts List©Copyright 2007, 4thEdition:Oct.,2011(YK)Subjecttochangewithoutnotice.Model ISC450G [Style: S2]I
CMPL12D06D05-02E24thEdition:Oct..2011(YK)2345311078963Pipe/Wall Mounting Hardware (Option code: /U)Awning Hood (Option code: /H5)Panel Mounting
Revision RecordManualTitle: ModelISC450G[Style:S2]InductiveConductivityConverter Manual Number : IM 12D06D05-01E Edition Date Remark (s)1s
IM 12D06D05-01Ev6. CALIBRATION ...366-1.Befor
1IM 12D06D05-01E1 INTRODUCTIONANDGENERALDESCRIPTION1. INTRODUCTION AND GENERAL DESCRIPTIONTheYokogawaEXAxtISC450Gisaconverterdesigned for i