Yokogawa PC-Based MX100 User Manual Page 82

  • Download
  • Add to my manuals
  • Print
  • Page
    / 133
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 81
3-10
IM MX100-01E
For -NDI
Connect CH2 only
Switch the
bridge head
R2
R4R1
R3
ON
1 2 3 4 5
1
2
3
4
5
6
7
8
OFF
SW
Bridge head
(701955 or 701956)
SW1
OFF
SW2
OFF
SW3
OFF
SW4
ON
SW5
OFF
When calibrating the output range of the 8-CH, Medium-Speed Analog Output module
All channels are calibrated at Zero (0 V) and Full (10 V).
Digital multimeter
YOKOGAWA 7562
H
L
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
3.3 Calibration