Yokogawa Button Operated MV2000 User Manual Page 57

  • Download
  • Add to my manuals
  • Print
  • Page
    / 310
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 56
1-48
IM MV1000-01E
1.10 FAIL/Status Output Function (/F1 Option)
FAIL Output
When a failure occurs in the MV CPU, the MV outputs a relay contact signal (1 signal).
The relay is energized when the CPU is normal and de-energizes when a CPU failure
occurs. Therefore, relay output is carried out when the power is turned OFF (or when
there is a power failure). You cannot change this behavior.
Relay Operation
NO C NC
During normal operation
NO C NC
When a failure occurs
NO C NC
When power is turned OFF
* NO: Normally Opened, C: Common, NC: Normally Closed
If a failure occurs, contact your nearest YOKOGAWA dealer for repairs.
Page view 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 309 310

Comments to this Manuals

No comments