<2. Specications>
2-21
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
2.7 Other Equipments
2.7.1 Dust Filter for the Detector (K9471UA)
This lter is used to protect the detector sensor from a corrosive dust components or high velocity
dust in recovery boilers and cement kilns.
Sample gas ow rate is needed to be 1 m/sec or more to replace gas inside zirconia sensor.
Standard specication
Applicable detector: Standard-type detector for general use (the sample gasow should be
approximately perpendicular to the probe.)
Mesh: 30 microns
Material: Carborundum (Filter), SUS316 (JIS)
Weight: Approx. 0.2 kg
Part No. Description
K9471UA Filter
K9471UX Tool
Ø51
32
10
Carborundum filter (SiC)
Detector
Screw
Increasing of insertion length
F3-2E.ai
Attach the filter unit to the tip of
the detector by screwing it clockwise.
2.7.2 Dust Guard Protector (K9471UC)
Recommended to be used when sample gas is likely to ow directly into the cell due to its ow
direction in the stack or the like, ammable dust may go into the cell, or water drops are likely to
fall and remain in the cell during downtime or the like due to the installation position.
Material: SUS316
Weight: Approx. 0.3 kg
4-Ø6
122
Ø50.8
100
Increasing of insertion length
F11-1E.ai